Specifications:
Main unit
■ Working modes: For the closed system, half closed system and half open system
■ Application: Adult and Pediatric
■ Low gas supply alarm: when O2 supply pressure is over low,
the machine will arise > (longer than) 7 seconds audible alarm.
■ High pressure limitation: when airway pressure is higher than pre-set figure, the machine will automatically start protection function. The airway pressure will stop increasing (not increase any more) to keep patient safe.
■ Low pressure protection: when O2 gas supply pressure is lower than 0.1MPa,the machine will automatically cut off N2O supply.
■ When power supply fails, audible and visual alarms will arise. The machine will continuously work by battery.
■ Respiratory loops are available.
■ O2 Flush: 25L/min~75L/min
■ Gas Supply: O2: 0.28MPa~0.6MPa N2O: 0.28MPa~0.6MPa Air: 0.28MPa~0.6MPa
■ Power supply: AC 220V, 50 Hz Built-in UPS power supply for backup.
Flow Meter
■ O2/N2O/Air six-tube Flowmeter (0~1L/min 1~10L/min )
■ Linkage of O2 and N2O, keep oxygen concentration over 25%
Anesthesia Ventilator
■ Bellow for Adult 1 piece Bellow for Pediatric 1 piece
■ Pneumatically driven and electrically control
■ Monitoring: under Manual control or Automatic control, Tidal volume(using differential pressure flow sensor), Frequency, I/E ratio; Peak airway pressure,
Wavefrom(Pressure-time),Wavefrom(Flow-time)
■ Respiratory modes: IPPV, SIGH, SIMV/f/2, SIPPV, PEEP(0.5~2kPa)
■ Tidal volume:0,50~1500mL( with tidal volume preset function)
■ Frequency: 2~99bpm
■ I:E:5:1~1:8 (continuously adjustable)
■ PTRIG: -2~0kPa
■ High pressure limitation:2~6kPa
■ Low pressure limitation:0~2kPa
Circle Absorber Assembly
■ New launching integrated CO2 Absorber, could be high temperature 134℃sterilized.
Cup type Canister, around 2.5L, easily and fast change soda lime.
Anesthesia Vaporizer
■ High precise Anesthesia Vaporizer (Enflurane, Halothane, Isoflurane, Sevoflurane)
■ Sevoflurane adjustment scope: 0~8% Vol
Enflurane, Halothane, Isoflurane adjustment scope:0~5%Vol
■ Automatic compensation of flow, temperature and pressure
■ Self-lock and interlock
■ Volume: ≥250mL