Technical Parameters
Main Unit:
Working modes: For the closed system, half-closed system and half-open system
Application: Adult and Pediatric
Low gas supply alarm: when Oz supply pressure is over low, the machine will arise > (longer than) 7 seconds audible alarm.
High pressure limitation: when airway pressure is higher than pre-set figure, the machine will automatically start protection function. The airway pressure will stop increasing (not increase any more) to keep patient safe.
Low pressure protection: when Oz gas supply pressure is lower than 0.1MPa,the machine will automatically cut off N20 supply.
When power supply fails, audible and visual alarms will arise. The machine will continuously work by battery.
O2 Flush: 25L/min~75L/min
Gas Supply: O2: 0.28MPa~0.6MPa N20: 0.28MPa~0.6MPa
Power supply: AC 220V, 50 Hz Built-in UPS power supply for backup.
Flowmeter:
02/N20 four-tube Flowmeter (0 ~ 1L/min 1 ~ 10L/min)
Linkage of Oz and N20, keep oxygen concentration over 25%
Anesthesia Ventilator:
Pneumatically driven and electrically control
Monitoring: Tidal volume, Frequency, MV, I/E ratio; Peak airway pressure, Oz concentration, Apnea. Waveform(Pressure-time), Waveform (Flow-time)
Working mode: Manual control or Automatic control
Respiratory modes: IPPV, SIGH, SIMV/f/2, SIPPV, PEEP(0.5kPa ~2 kPa) Optional: EtCOz
Tidal volume:0,50 ~ 1500mL
Respiratory frequency:2 ~ 99bpm
I:E:5:1 ~ 1:8 adjustable
MV: not lower than 16L/min
FiOz: 21 ~ 100%
PtriG: -2 ~ 0kPa
High pressure limitation:2 ~ 6kPa Low pressure limitation:0 ~ 2kPa
Circle Absorber:
New launching integrated COz Absorber, could be high temperature 134°C sterilized.
Cup type Canister, around 2.5L, easily and fast change soda lime.
Anesthesia Vaporizer:
High precise Anesthesia Vaporizer (Enflurane, Halothane, Isoflurane, Sevoflurane)
Sevoflurane adjustment scope: 0 ~ 8% Vol;
Enflurane, Halothane, Isoflurane adjustment scope: 0 ~ 5%Vol
Automatic compensation of flow, temperature and pressure
Self-lock and interlock
Volume: ≥250mL