SMT-205 Anesthesia Machine

  • Parameters

  • Parameters

Technical Parameters

Main Unit:

Working modes: For the closed system, half-closed system and half-open system

Application: Adult and Pediatric

Low gas supply alarm: when Oz supply pressure is over low, the machine will arise > (longer than) 7 seconds audible alarm.

High pressure limitation: when airway pressure is higher than pre-set figure, the machine will automatically start protection function. The airway pressure will stop increasing (not increase any more) to keep patient safe.

Low pressure protection: when Oz gas supply pressure is lower than 0.1MPa,the machine will automatically cut off N20 supply.

When power supply fails, audible and visual alarms will arise. The machine will continuously work by battery.

O2 Flush: 25L/min~75L/min

Gas Supply: O2: 0.28MPa~0.6MPa N20: 0.28MPa~0.6MPa

Power supply: AC 220V, 50 Hz Built-in UPS power supply for backup.

Flowmeter:

02/N20 four-tube Flowmeter (0 ~ 1L/min 1 ~ 10L/min)

Linkage of Oz and N20, keep oxygen concentration over 25%


Anesthesia Ventilator:

Pneumatically driven and electrically control

Monitoring: Tidal volume, Frequency, MV, I/E ratio; Peak airway pressure, Oz concentration, Apnea. Waveform(Pressure-time), Waveform (Flow-time)

Working mode: Manual control or Automatic control

Respiratory modes: IPPV, SIGH, SIMV/f/2, SIPPV, PEEP(0.5kPa ~2 kPa) Optional: EtCOz

Tidal volume:0,50 ~ 1500mL

Respiratory frequency:2 ~ 99bpm

I:E:5:1 ~ 1:8 adjustable

MV: not lower than 16L/min

FiOz: 21 ~ 100%

PtriG: -2 ~ 0kPa

High pressure limitation:2 ~ 6kPa Low pressure limitation:0 ~ 2kPa
Circle Absorber:

New launching integrated COz Absorber, could be high temperature 134°C sterilized.

Cup type Canister, around 2.5L, easily and fast change soda lime.

 

 


Anesthesia Vaporizer:

High precise Anesthesia Vaporizer (Enflurane, Halothane, Isoflurane, Sevoflurane)

Sevoflurane adjustment scope: 0 ~ 8% Vol;
Enflurane, Halothane, Isoflurane adjustment scope: 0 ~ 5%Vol

Automatic compensation of flow, temperature and pressure

Self-lock and interlock

Volume: ≥250mL

 

Related Products

SMT-30F Medical Ventilator

High-flow nasal cannula SMT-30F is a High Flow Nasal Cannula Oxygen Therapy Device capable of delivering up to 100% humidified and heated oxygen at a flow rate of up to 80L/min, it delivers a mixed oxygen with an easy controlled temperature and humidity. The oxygen supply system provides a great solution of airway management and oxygenation improvement.

Leave A Message


    Products
    Contact
    WhatsApp
    Email