Specification:
Main unit
■ Working modes: For the closed system, half closed system and half open system
■ Application: Adult and Child and Neonate
■ Low gas supply alarm: when O2 supply pressure is over low, the machine will
arise > (longer than) 7 seconds audible alarm.
■ High pressure limitation: when airway pressure is higher than pre-set figure,
the machine will automatically start protection function. The airway pressure
will stop increasing (not increase any more) to keep patient safe.
■ Low pressure protection: when O2 gas supply pressure is lower than 0.1MPa,
the machine will automatically cut off N2O supply.
■ When power supply fails, audible and visual alarms will arise. The machine will continuously work by battery.
■ O2 Flush: 25L/min~75L/min
■ Gas Supply: O2: 0.28MPa~0.6MPa N2O: 0.28MPa~0.6MPa Air: 0.28MPa~0.6MPa
■ Power supply: AC 220V, 50 Hz Built-in large volume Lithium battery, could continuously work for 3 hours.
Flow Meter
■ O2/N2O/Air six-tube Flowmeter (0~1L/min 1~10L/min )
■ Linkage of O2 and N2O, keep oxygen concentration over 25%
Anesthesia Ventilator
■ Pneumatically driven and electronically controlled
■ Monitoring: VTI, VTE,Frequency, MV, I:E, Ppeak, Pmean, Pplat, Pmin, PEEP, FiO2, Cst, Cdyn, Rst Optional: Et CO2
■ Respiratory modes: VCV, PCV, PSV, SIMV-VC, SIMV-PC, PRVC, SIMV-PRVC
■Tidal volume:0,20~1500mL Tslope: 0~2s
■Frequency: 1~100bpm Esens: 5~80%
■I:E:4:1~1:10 Fsens: 0.5~20L/min.
■Tinsp: 0.1~10s Psens: 0~20cmH2O
■Pause: 0~50% Pressure support: 0~70cmH2O
■PEEP: OFF, 3~30cmH2O Pressure control: 5~70cmH2O
Electronic PEEP, convenient and exact. The variable orifice type flow sensor from Germany, makes sure the flow rate exactly be detected.
■By intelligent alarm system, the alarm limits could be preset, and alarm logs(200 pieces)could be recorded.
Circle Absorber Assembly
■ New launching integrated CO2 Absorber, could be high temperature 134℃sterilized.
Cup type Canister, around 2.5L, easily and fast change soda lime.
Anesthesia Vaporizer
■ High precise Anesthesia Vaporizer (Enflurane, Halothane, Isoflurane, Sevoflurane)
■ Sevoflurane adjustment scope: 0~8% Vol
Enflurane, Halothane, Isoflurane adjustment scope:0~5%Vol
■ Automatic compensation of flow, temperature and pressure
■ Self-lock and interlock
■ Volume: ≥250mL