Specification:
Main unit
■ Working mode: For the closed system, half-closed system and half–open system
■ Application: Adult, Pediatric
■ Low gas supply alarm: when O2 supply pressure is over low,
the machine will arise >(longer than) 7 seconds audible alarm.
■ High pressure limitation: when airway pressure is higher than pre-set figure, the machine will automatically start protection function. The airway pressure will stop increasing (not increase any more) to keep patient safe.
■ Low pressure protection: when O2 gas supply pressure is lower than 0.1MPa, the machine will automatically cut off N2O supply.
■ When power supply fails, audible and visual alarms will arise.
■ O2 Flush: 25L/min~75L/min
■ Gas Supply: O2: 0.28MPa~0.6MPa N2O: 0.28MPa~0.6MPa
■ Power supply: AC 220V, 50 Hz
Flowmeter
■O2 / N2O two tubes Flowmeter (0~10L/min)
■ O2 / N2O linkage system, make sure the O2 concentration > 25%
Anesthesia Ventilator
■ Bellow for Adult 1 piece
■ Pneumatically driven and electrically controlled
■ Monitoring: Tidal volume, Frequency; I/E Ratio, Peak airway pressure, Waveform (P-T)
■ Respiratory modes: Manual, IPPV, SIGH,SIMV/F/2,SIPPV
■ Tidal volume:0,50~1500mL
■ Respiratory frequency:2~99bpm
■ I:E:5:1~1:8 (adjustable)
■ PTRIG: -2~0kPa
■ High pressure limitation:2~6kPa
■ Low pressure limitation:0~2kPa
Circle Absorber Assembly
■ New launching integrated CO2 Absorber, could be high temperature 134℃sterilized.
Cup type Canister, around 2.5L, easily and fast change soda lime.